Hello! Welcome to Jiangsu Zhongwei Measurement and Control Instrument Co., Ltd.

Seismic type vortex flowmeter

5抗震型涡街流量计.png
  • 5抗震型涡街流量计.png
Seismic type vortex flowmeter is mainly used for industrial piping medium fluid flow measurement, su…

Description:

Products overview:
ZW - LUGB series vortex flowmeter are resistantIs based on the principle of karman vortex street and successfully developed a kind of advanced international type fluid turbulence flow meter.Because it has the advantages of other flowmeter it too, since the seventy s got rapid development, widely used in petroleum, chemical industry, metallurgy, paper making, power, water conservancy and other industries the measurement of fluid.itWith its micro power consumption can be convenient access control system and computer system, to meet customer requirements under different working condition, it is not affected by medium, the influence of temperature, pressure, viscosity and component without plugging, BuKa, not easy to scale, can also be resistant to high temperature and high pressure, safety explosion-proof, suitable for bad environment.Introduced seismic intelligent vortex flowmeter in particular, it has excellent seismic performance, can automatically identify the vibration signal and traffic signal, at the same time the output seismic processing and non seismic processing the pulse signal.The flowmeter have no moving parts, high reliability, high precision, long life, can be in a very wide range of flow accurately measure the instantaneous flow and cumulative flow of the fluid.

Second, the working principle:
ZW - LUGB seriesSeismic type vortex flowmeter is by the design, testing probe in the flow field in the vortex body and the corresponding electronic circuit, etc.When the fluid flows through the spiral body occurs, it is formed on both sides of the two rows of vortex variations, this vortex is known as the karman vortex street.On the basis of the theory of karman vortex street and proposed the karman vortex street frequency is proportional to the velocity of the fluid, the relation between the frequency and velocity are presented: f = St * V/d type:
F vortex frequency (Hz)
V vortex occurred body on both sides of the average flow velocity (m/s)
St, luo hall coefficient (constant)
These variations of vortex is formed a series of variations of negative pressure, the pressure on detecting probe, and produce a series of alternating electric signals, after dealing with the preamplifier conversion, plastic, amplification, the output is proportional to the vortex synchronous pulse frequency signal (or standard).
Applications:
ZW - LUGB seriesSeismic type vortex flowmeter has the range than wide, high precision, easy installation and maintenance and a series of advantages such as wide adaptability of medium.Can be widely used in petroleum, chemical industry, metallurgy, machinery, food, papermaking, and urban pipeline heating, water supply, gas and other industries of all kinds of low viscosity liquid, gas, steam, etc of single-phase fluid process measurement and energy saving management.
Four, the main features:
1, can accurately measure within wide flow range of gas, liquid and steam flow and not affected by fluid physical properties;
2, not affected by temperature, pressure, not easy to plug at the same time, not easy card, not easy to scale, high temperature resistant, high voltage;
3, safety explosion-proof, suitable for bad environment;
4, have no moving parts, no hole aperture design, product without any abrasion, corrosion resistance, dirt without mechanical maintenance, long service life;
5, using micro power consumption of new and high technology, a battery-powered scene shows type flowmeter, power is not running for more than two years;
6, temperature, pressure compensation integrated design;
7, current output are electrical isolation, good common-mode interference suppression ability;
8, also showed flow values and cumulative flow value, don't have to take turns to switch;
9, using vibration probe, the effective elimination of the influence of external vibration;
10, using split type signal converters, cable zui, 100 meters long;
11, than a wide range of those days;
12, instrument structure design is reasonable, wide dynamic range, small pressure loss;
13, table body is made of stainless steel material, can be applied to measurements of corrosive medium;
14, the liquid crystal display, pulse, 4 ~ 20 ma output 485 newsletter, or can be connected to the industrial automation system.
Five, the technical parameters:
1, the measured medium: gas, liquid and vapor;
2, connection mode: flange card mounted, flange, plug-in;
3, diameter specifications flange mounted caliber select 25,32,50,80,100;
4, the flange connection type choose 100150200;
5, flow range (see chart flow range)
6, normal measuring velocity range Reynolds number 104 ~ 1.5 x 4 x 106;Gas 5 to 50 m/s;Liquid 0.5 ~ 7 m/s;
7, measurement accuracy 1.0 magnitude 1.5 on the Richter scale;
8, measured medium temperature: room temperature - 25 ℃ ~ 100 ℃, high temperature - 25 ℃ ~ 150 ℃ to 25 ℃ ~ 250 ℃;
9, high output signal pulse voltage output signal level 8 ~ 10 V low level from 0.7 V to 1.3 V;
10, pulse duty ratio is about 50%, transmission distance is 100 m;
11, pulse current 4 ~ 20 mA remote transmission signal, transmission distance is 1000 m;
12, instrument: using the environment temperature 25 ℃ ~ + 55 ℃, humidity: 5 ~ 90% RH50 ℃;
13, the material of stainless steel, aluminum alloy;
14, DC24V power supply or lithium battery 3.6 V;
15, explosion-proof grade this Ann model iaIIbT3 - T6, protective class IP65.

Six, appearance size:

DN

A

B

C

D

15

90

Φ 57

383

45

20

100

Φ 57

388

50

25

100

Φ 57

394

50

32

100

Φ 65

396

50

40

100

Φ 75

401

50

50

110

Φ 87

407

55

65

110

Φ 109

418

55

80

110

Φ 120

423

55

100

120

Φ 149

447

60

125

125

Φ 175

474

65

150

145

Φ 203

501

75

200

170

Φ 259

556

100

250

190

Φ 312

608

120

300

210

Φ 363

660

140

350

230

Φ 409

709

160

400

250

Φ 460

756

180

450

275

Φ 520

814

205

500

290

Φ 575

869

225

Seven, pipe requirements:
1, the requirement for straight pipe:
In order to ensure that the instrument, accurate to run normally, sensor must have a certain straight pipe upstream and downstream of the installation points, to adjust the flow field, as shown.

Figure 1: concentric contraction tube;
Figure 2: the concentric expansion tube;
Figure 3: a 90 - degree Angle;
Figure 4: the same plane of two 90 - degree Angle;
Figure 5: different plane of two 90 - degree Angle;
Figure 6: regulating valve should be installed downstream of the sensor in 5 d to the distance, if the sensor must be installed in the upstream, the sensors should be not less than 50 d upstream, straight pipe section downstream straight pipe section should be not less than 5 d.
2, the requirement of pipe:
A, upstream and downstream piping diameter D and the same as the sensor inner diameter DN, its difference satisfy the following conditions: 0.95 DN D 1.1 or less or less DN.
B, piping shall be concentric with sensor, coaxial degree should be less than 0.05 DN.
C, gasket not protruding into the pipe, the inner diameter slightly larger than sensors.
D, if you want to stop inspection and cleaning the sensor, the by-pass pipe should be set up,As shown in the figure below.

 
3, the requirement of pipe vibration:
Sensors to avoid installed on the strong pipeline vibration, if necessary to install, shock absorption measures must be taken, respectively at the upstream and downstream of sensor 2 d piping fastening device, and vibration-proof pad.
Special note: in the air compressor exit strong vibration, cannot install the sensors, shall be installed after the gas storage.
4, demand for air condition:
A, the gas to be tested for single phase pipe flow, continuously flowing through a pipe.
Before b, the gas flows through the flowmeter, the flow rate must be parallel to the pipe axis, can not have vortex flow.
C, flow should be a subsonic, a pulse, its flow change slowly over time.
5, the flowmeter installation requirements:
Vortex street flowmeter to process the installation and use of the special requirements of environment is not too much, but any kind of flow measurement instrument has such a common, as far as possible to avoid vibration and high temperature environment as flow interference elements (such as compressor, separator, pressure regulating valve, reducer, and pipe, elbow, etc.), keep straight pipe before and after the instrument inside sleek straight, ensure to clean single-phase fluid measured medium, etc.

Eight, selection of key points:
Correct selection to ensure that the use of vortex flowmeter is better.Should choose what kind of vortex street flowmeter is based on the current body to determine the physical properties and chemical properties of the medium.Make the size of vortex flowmeter, flow range, lining materials, electrode materials and output current, etc.To be the nature of the flow volume and flow measurement requirements.
1, precision function
Precision grade and function in accordance with the requirements of measurement and choose the instrument using fine grade, cost-effective.Such as for trade settlement, product transfer and energy measurement occasions, should choose high precision grade, such as 1.0, 0.5, or higher level;Used in process control, according to the requirements of control choose not with precision grade;Some just check process flow, without having to do precise control and measurement, can choose lower precision grade, such as 1.5, 2.5, 4.0, and even then you can choose cheap plug-in vortex flowmeter.
2, medium can be measured
Medium velocity and diameter, full scale measurement generally medium, the full flow of vortex flowmeter can measure medium velocity in 0.5-12 m/s within the scope of selection, range is wide.Choose different instrument specifications (diameter) and process piping, the same shall consider whether measuring flow range within the scope of the flow velocity, namely when the pipeline flow rate is low, can not meet the requirement of the flow meter, or under the velocity measurement accuracy of the quasi cannot guarantee, the need to reduce the size instrument, which carry high tube flow velocity, measurement results are satisfied.
3, the selection of vortex flow transmitter
In saturated vapor measurement using VA type piezoelectric vortex flow transmitter, due to the vortex flow meter range wide, therefore, in practice, the measurement of saturated steam flow main consideration shall not be less than the lower limit of vortex flowmeters, that is to say, must satisfy the fluid flow rate shall be not less than 5 m/s.According to the size of the YongQiLiang chooses different caliber of vortex flow transmitter, rather than with the existing technology to choose transmitter diameter pipe size.
The choice of 4 pressure transmitter, pressure compensation
Because of the saturated steam pipe length, pressure fluctuation is bigger, pressure compensation must be used, considering the pressure, temperature and density, the relation between measurement using pressure compensation, because the pipeline saturated steam pressure in the range 0.3 0.7 MPa, pressure transmitter ranges 1 MPa.
5, display instrument selection
Display instrument intelligent flow display, a voltage compensation, the instantaneous flow and cumulative flow integrating function.
Nine, field requirements:
1, low maintenance - most of the vortex flowmeter are on the market with induction swirl hole or plug-in measurement, once the impurities embedded in the medium pressure hole or sensing element and the table body, the gap between the signal is weak or unstable.A good design should be not easy to plug part, thereby reducing amount of maintenance.
2, inductive components online replacement - some manufacturer of inductive components and vortex street close 2 for one, seemingly simple design to the actual use of inconvenience.Because once induction part failure, you need to medium to the pipe emptying replacement parts, after pressure relief effect production.Perfect design should be separated the two, so can replace the induction part alone, without emptying the media.
3, vortex flowmeter is vulnerable to vibration interference, well-designed vortex flowmeter can through hardware and digital signal processing to exclude interference, and the stability of the signal is obtained.
4, vortex flow meter before and after the installation of a hassle to very long straight section, some manufacturers can offer within the flowmeter reducing design, greatly reduces the user specifically MenWei vortex flowmeter with the demand of the straight pipe.
Ten problems need to pay attention to during using, debugging:
1, flowmeter installation, maintenance, must abide by the "do not open your have explosion-proof gas" warnings, and turned off before open external power.
2, vortex flow meter used in outdoor, suggestion and side shields.
3, shall not be arbitrarily fixed part loosen the flowmeter.The protection of the pressure sensor.Users in the use of traffic time, should pay attention to the flow pressure overload value of pressure sensor is rated working pressure of 1.5 times (that is, the media zui high pressure of 1.5 times).Selection should be within the prescribed flow range, prevent overspeed operation, in order to obtain the ideal accuracy and ensure the normal service life.(due to pressure test, purging pipes or exhaust overdrive, and turbine operation in the reverse flow meter can cause damage)
4, when put into operation, should slowly open valve to prevent damage of instantaneous flow and the turbine.Add lubricant should be according to the sign, the number of refueling depends on temperament clean degree, usually 2 ~ 3 times a year.
5, runtime or alternate internal related parameters are not allowed to open the rear cover, otherwise will affect the normal operation of the meter.
Eleven, install and use should be paid attention to in the items:
Vortex street no mechanical moving parts, small size, corrosion resistance, stable performance, can be directly display medium work pressure, temperature, flow rate and volume under standard state, wide range, the advantages of small measurement error, has been widely used in measurement and gas metering of oil and gas well production.In years of field application, we think the flowmeter zui suitable for relatively clean dry gas metering, gradually become small and medium-sized gas metering.To guarantee the accuracy of the flowmeter measurement, reduce the measuring error, please note:
1, the selection of reasonable
Don't make the actual zui small flow meter QVmin below the selected through runoff flow range limit.
2, installation is correct
Upstream of the flow meter should be at least 10 d measuring straight pipe length.Straight pipe inner surface should be smooth, cannot have the steps, more smooth connecting flange weld, flange gasket can't into the pipe.Don't put the wrong installation direction of the flow meter.Tests showed that: all the unreasonable installation, will lead to the flowmeter measurement value is low.
3, away from the source of shocks, or install piers in upstream or downstream flow meter, in order to avoid the influence of the pipeline vibration.Avoid strong magnetic field disturbance to the flow meter chip.
4, temperament is relatively clean
Especially not a gas liquid two phase flow;Temperament is dirty, should clean instant flow meter.
5, and reduce the influence of flow pulsation
It can be installed in front of the flow meter buffer tank, or extended length of straight pipe upstream of the flow meter, computation.
6, timely flowmeter for metrological verification, to ensure accurate and reliable value.

Twelve, installation requirements:
1, the sensor can be installed indoors, can also be installed in the outdoor.Environmental conditions to conform to the requirements.
2, sensors should be installed in a horizontal, vertical or inclined (fluid flow from bottom to top) and its corresponding pipe nominal diameter.
3, sensors installed on a mechanical vibration of pipeline should be avoided.Should be considered when the vibration is inevitable, at about 2 dn from sensor before and after the straight pipe and fixing racks.
4, avoid install the sensor sensing has strong electromagnetic interference, thermal radiation, corrosive gas, the space is little, and maintenance is not convenient.
5, when measured medium contains many impurities, at the request of straight pipe upstream from sensors mounted outside the length of the filter.
6, either upstream or downstream of the sensor should be configured with a certain length of straight pipe, straight pipe inner surface should be clean, smooth, no obvious convex concave, fouling and peeling phenomenon.Its length should conform to the requirements of figure 2.Installation of the liquid sensor, near the pipe should be filled with liquid to be tested.
7, straight pipe diameter as far as possible consistent with the sensor size, if not consistent, slightly larger diameter than the sensor size should be adopted, the error to 3% or less is not more than 5 mm.
13, the matters needing attention:
1, special flange and straight pipe welding can't with the sensor.
2, when installation should make the flow of sensor mark the direction of flow and the pipe.
3, before the installation of sensor and the flange must be put away in sealing ring grooves.The location of the pressure and temperature measurement point, pressure downstream of the sensor in 3 ~ 5 dn, temperature measuring points in the downstream of 5 ~ 8 dn.
4, when measuring high temperature medium, do not use heat insulation material to wrap around the sensor link rod.
Five, connection of shielded cable, should as far as possible away from the interference of strong electromagnetic field.Absolutely not allowed with high voltage cable laying, shielded cable to shorten as far as possible, and shall not be volume, in order to reduce stray inductance, length should not exceed 200 m.
6, before installation of sensors, pipeline cleaning must be conducted.Flow after rinse tube impurities, avoid clogging sensor.Measuring liquid pipe must be full of measured liquid, prevent the interference of bubbles.
7, measuring gas pipelines to prevent the interference of reservoir fluid.Replace the probe body under the high temperature and high pressure, it must be safe operation, high temperature protection.After cooling step-down replaceable probe at the bottom of the safety conditions.

14, standard of flow range profile: unit (m3 / h)

traffic 

The liquid flow range

Gas flow range

Saturated steam flow range

CAL

Small flow

Heavy traffic

Small flow

Heavy traffic

Small flow

Heavy traffic

DN25

0.9

14

8.3

110

13

1500

DN40

2.5

35

26

300

25

4400

DN50

3.3

55

36

480

40

6800

DN80

8

150

90

1300

100

19000

DN100

14

240

140

2000

160

29000

DN150

38

450

290

4100

350

66000

DN200

70

850

620

7500

620

118000

DN250

130

1300

700

12500

970

185000

DN300

180

2000

920

16500

1400

267000

15, selection of coding:

Code name

size

Flow range ㎡/h(In the wei and control)

ZW-LUGB-25

DN25

1 ~ 10(liquid)

25 ~ 60(gas)

Steam flow check specifications,DN300The above recommended plug-in vortex flowmeter

ZW-LUGB-32

DN32

1.5 ~ 18(liquid)

15-150(gas)

ZW-LUGB-40

DN40

2.2 ~ 27(liquid)

22.6 ~ 150(gas)

ZW-LUGB-50

DN50

4 ~ 55(liquid)

35 ~ 350(gas)

ZW-LUGB-80

DN80

9 ~ 135(liquid)

90 ~ 900(gas)

ZW-LUGB-100

DN100

14 ~ 200(liquid)

140 ~ 1400(gas)

ZW-LUGB-150

DN150

32 ~ 480(liquid)

300 ~ 3000(gas)

ZW-LUGB-200

DN200

56-800(liquid)

550 ~ 5500(gas)

Code name

function1

N

No temperature pressure compensation

Y

With temperature compensation

Code name

Output type

F1

4-20mAOutput (two-wire system)

F2

4-20mAOutput (three-wire system)

F3

RS485Communication interface

Code name

Measured medium

J1

liquid

J2

gas

J3

steam

Code name

The connection method

L1

Flange mounted

L2

Flange connection type

Code name

function2

E1

1.0level

E2

1.5level

T1

The normal temperature

T2

The high temperature

T3

steam

P1

1.6MPa

P2

2.5MPa

P3

4.0MPa

W1

internal3.6VThe power supply

W2

DC24VThe power supply

N1

Stainless steel

N2

Carbon steel

N3

Anti-corrosion type

16,Order to choose selection guidelines:

Name and model number of 1, 1), (2) the diameter, flow range, (3) whether or not to bring the attached so that we can correct selection for you, (4) using pressure, (5) the use of medium temperature.

2, if has been selected by design units in the way of vortex street flowmeter types, please press type vortex flowmeter directly order from our sales department.

3, when using occasions is very important, or the environment is complex, please try to provide the design drawings and detailed parameters, by wei experts in our audit checks for you.

After-sales service commitment:

1, from the date of signing of the contract, I in the way provided for products of the company offers free maintenance and maintenance service, commitment to lifelong maintenance service;

2, jiangsu wei measurement and control instrument co., LTD. Sales manager will be in regular contact with customers, understand the usage of products, and solve the problems of the customers to use in the process of produce, free of charge to provide technical support;

3 the man-made damage is found, the warranty period, our company is responsible for the maintenance, and the resulting maintenance fee;

4, product quality problems or are not satisfied with product, the user can choose unconditional return, wei tt&c don't charge any fees, appear quality problem, bear the freight back and forth.


Browse:
Return To The Top

分享到: